27 ProtocolWatch
354
Checking The Sepsis Resuscitation Bundle Window
During the Sepsis Resuscitation Bundle you can open the Sepsis Resuscitation Bundle window at any
time to check the status or to make entries. The recommendations are shown as follows:
Recommendations which are not yet fulfilled are shown in black. They also appear in the list on
the lower area of the
ProtocolWatch Screen.
Recommendations which are overdue are shown in orange.
Recommendations which have been fulfilled and confirmed are shown in gray. They no longer
appear in the list on the lower area of the
ProtocolWatch Screen.
Previewing the Sepsis Management Bundle
At any time during the Sepsis Resuscitation Bundle you can preview the Management Bundle in order
to see which recommendations will follow there.
Select the
Preview Managmnt key.
Ending The Sepsis Resuscitation Bundle
When all recommendations have been fulfilled and confirmed, or when 6 hours have passed since the
bundle started, the Sepsis Resuscitation Bundle is ended.
You are asked to confirm with the
Yes key that the Sepsis Management Bundle should now be started.
If you select
No, the Sepsis Resuscitation Bundle is extended for 1 hour.
Sepsis Management Bundle
The stabilization recommendations from the Sepsis Management Bundle are presented and the
implementation of the recommendations is monitored.
When the Sepsis Management Bundle phase begins, the
ProtocolWatch Screen is automatically
updated (if configured and available on the monitor). In the lower area of the Screen, the list of
recommendations for the Sepsis Management Bundle replaces the list for the Sepsis Resuscitation
Bundle. The timer shows the combined time for both phases.
Monitoring the Sepsis Management Bundle Recommendations
When you select the lower area of the ProtocolWatch Screen, the Sepsis Management Bundle
window appears. If you do not have the
ProtocolWatch Screen displayed, the Sepsis Management
Bundle
window appears automatically once every six hours.
The
Low Dose Steroids for septic shock per policy recommendation may not appear if it has been
configured not to do so for your hospital.
Check the
Low Dose Steroids administered box when you have started steroid therapy or the Low
Dose Steroids contra-indicated
box if this treatment is contra-indicated for your patient.
The boxes for
Maintain Glucose <8.3mmol/l (150 mg/dl) and Maintain Insp. Plateau Pressure
<=30cmHO
can only be checked when the timer has reached 20 hours to ensure that the stabilization
of these values is achieved long-term. Check these boxes if the recommendation has been fulfilled.
Recommendations which have been fulfilled are shown in gray with the time fulfillment was entered.
To see more details about the recommendations, select the
Show Details key.
Select
Confirm when the check boxes reflect the current treatment status.