MX600/700/800 Performance Specifications
Alarm Signal System alarm delay
The system alarm delay is the
processing time the system needs
for any alarm to be indicated on the
monitor, after the measurement has
triggered the alarm.
less than 4 seconds
Delay for alarm availability on the
network
This is the time needed after alarm
indication on the monitor until the
alarm signal is available on the
network, to the IntelliVue
Information Center or for
transmission to other systems.
less than 5 seconds
Pause duration 1,2,3 minutes or infinite, depending on configuration
Extended alarm pause 5 or 10 minutes
Sound pressure range min. 0 dB(A)
max. 45-85 dB(A)
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity 300 items
Real Time Clock Range from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy better than 4 seconds per day
Hold Time infinite if powered by AC; otherwise at least 24 hours
Buffered Memory Hold Time if powered by AC: infinite
without power: at least 8 hours
Contents Active settings, trends, patient data, realtime reports, events,
review alarms
MX600/700/800
Power Specifications Power Consumption <200 W average
Line Voltage 100 to 240 V
Current 1.9 to 0.9 A
Frequency 50/60 Hz
Indicators Alarms Off red or yellow LED with crossed out alarms symbol
Alarms red/yellow/light blue (cyan) LED
On/Standby/Error green/red LED integrated in power switch
External Power green LED
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO
2
modulation tone
4 different alarm sounds
Remote tone for alarms on other beds in network
Tone for Timer expired