40 Specifications
465
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO
2
modulation tone
4 different alarm sounds
Remote tone for alarms on other beds in network
Tone for Timer expired
Display Wave Speeds Available for standard waves 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
with ±5% accuracy (guaranteed only for integrated displays)
Available for EEG and BIS waves 6.25 mm/s, 12.5 mm/s, 15 mm/s, 25 mm/s, 30 mm/s, 50
mm/s
with ±5% accuracy (guaranteed only for integrated displays)
Trends Resolution 12, 16, 24 or 32 numerics @ 12 sec, 1 minute, 5 minute
resolution
Information Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
neonatal extended 12 numerics, 24 hours @ 12 secs or
32 numerics 32 hours @ 1 minute
intensive care extended: 16 numerics 120 hours @ 5 minutes
anesthesia extended 32 numerics 9 hours @ 12 seconds
High Res Trend Waves Measurements available HR, SpO
2
, Resp, tcpO
2
, Pulse, Perf, tcpCO
2
, CO
2
, A BP, PA P,
CVP, ICP, CPP, BIS, CCO, AWP, Anesthetic Agents, Delta
SpO
2
, inO
2
Resolution Measurement samples are taken at a resolution of four samples
per second
Update speed Waves are drawn at a speed of 3 cm/minute
Events Information Trigger condition and time, event classification and associated
detailed view of episode data
Episode data Configurable, either:
4 minutes of high resolution trend or
20 minutes of numerics trend @ 12 sec. resolution or
15 seconds of 4 waves @ 125 samples/sec. (Snapshot)
including all current numerics, alarms and inops
Capacity (max.) 25 or 50 events for either 8 or 24 hours
MX400/450/500/550