4 Patient Alarms and INOPs
84
Battery INOPs
Unsupported LAN
INOP tone
There is a problem with the communication to the network and central monitoring is currently not
possible. Check the connection. If the INOP persists, switch off the monitor and contact your service
personnel.
User I/F Malfunct
INOP tone
Perform a visual and functional check of all the monitor input devices. Contact your service personnel.
INOP Message, Indication What to do
INOP Message, Indication What to do
Batt 1 Missing
Batt 2 Missing
INOP tone
During this INOP, alarms cannot
be paused or switched off.
The monitor requires two batteries but can detect only one battery. Insert the missing battery
immediately.
Batt Empty
!!Batt Empty
!!!Batt Empty
INOP tone, battery LED flashes
During this INOP, alarms cannot
be paused or switched off.
The estimated remaining battery-powered operating time is less than 10 minutes. Replace the battery
immediately.
If the condition persists and the monitor is not connected to mains power, this INOP is re-issued two
minutes after you acknowledge it.
Batt Extensn Malf
INOP tone
There is a hardware error in the Battery Extension. Contact your service personnel.
Batt Incompat
INOP tone
The battery cannot be used with this monitor. Replace with the correct battery as specified in this
book.
Batt Low
INOP tone
The estimated battery-powered operating time remaining is less than 20 minutes.
Batt Malfunction
INOP tone, battery LED flashes
During this INOP, alarms cannot
be paused or switched off unless
the monitor is connected to
mains power.
The monitor cannot determine the battery status. If this INOP persists, replace the faulty battery. If
the condition persists and the monitor is not connected to mains power, this INOP is re-issued two
minutes after you acknowledge it.
Place the battery in a different monitor or in a battery charger. If the same INOP is shown, contact
your service personnel.
Batteries Empty
!!Batteries Empty
!!!Batteries Empty
Batt 1 Empty
!!Batt 1 Empty
!!!Batt 1 Empty
Batt 2 Empty
!!Batt 2 Empty
!!!Batt 2 Empty
INOP tone, battery LED flashes
During this INOP, alarms cannot
be paused or switched off.
The estimated remaining battery-powered operating time of the indicated battery or batteries is less
than 10 minutes. Replace the batteries immediately.
If the condition persists and the monitor is not connected to mains power, this INOP is re-issued two
minutes after you acknowledge it.
Batteries Incompat
Batt 1 Incompat
Batt 2 Incompat
INOP tone
The indicated battery or batteries cannot be used with this monitor. Replace with the correct battery or
batteries as specified in this book.
Batteries Low
Batt 1 Low
Batt 2 Low
INOP tone
The estimated battery-powered operating time remaining is less than 20 minutes.