6 ECG, Arrhythmia, ST and QT Monitoring
179
Switching Individual QTc Alarms On and Off
Each QTc alarm can be switched off individually.
To switch an alarm on or off, in the
Setup QT Analysis menu, select QTc High Alarm or ΔQTc High
Alarm
to toggle between On and Off.
Changing QTc Alarm Limits
Set the high alarm limits based on your assessment of the patient's clinical condition, unit protocols,
physician orders or medication specified limits.
1 In the Setup QT Analysis menu, select QTc High Limit.
2 Select the appropriate setting.
3 Select ΔQTc High Limit.
4 Select the appropriate setting.
Switching QT Monitoring On and Off
To switch all QT monitoring on or off, in the Setup QT Analysis menu, select QT Analysis to switch
between
On and Off.