41 Default Settings Appendix
512
Pulse Default Settings
1
When Pulse is not the alarm source, the alarms will be Off.
Pulse alarms use the settings of the currently selected Pulse alarm source.
STE Fem. V1,V4-6, Alarm Limit 1.0 mm 1.0 mm 1.0 mm
ST Settings - Lead I, II, III, V, aVR, aVL, V1-
6, MCL
Factory Defaults
Factory Adult Factory Pedi Factory Neo
QT Settings
Factory Defaults
Factory Adult Factory Pedi Factory Neo
QT Lead All All All
QTc High Limit 500 ms 480 ms 460 ms
ΔQTc High Limit 60 ms 60 ms 60 ms
QTc High Alarm OnOnOn
ΔQTc High Alarm OnOnOn
QT Analysis Off Off Off
QTc Formula Bazett Bazett Bazett
Pulse Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Alarm Source Auto Auto Auto Auto Auto Auto
Pulse (source label) On On On
System Pulse SpO₂ SpO₂ SpO₂ Auto Auto Auto
Alarms Off Enabled Enabled Enabled
AlarmSource Sel. Enabled Enabled Enabled
Pulse Alarm Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Pulse Alarms1 On On On
High Limit 120 bpm 160 bpm 200 bpm
Low Limit 50 bpm 75 bpm 100 bpm
Δ ExtrBrady 20 bpm 20 bpm 20 bpm
Brady Clamp 40 bpm40 bpm50 bpm40 bpm60 bpm80 bpm
Δ ExtrTachy 20 bpm 20 bpm 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm