40 Specifications
491
Cardiac Output / Continuous Cardiac Output
Complies with IEC 60601-2-34:2000
Negative Inspiratory
Pressure numeric
Range -20 to 0 cmH
2
O
Resolution 1 cmH
2
O
Mean Airway Pressure Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Plateau Pressure
numeric
Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Positive End-Expiratory
Pressure numeric
(PEEP)
Range 0 to 99 cmH
2
O
Resolution 1 cmH
2
O
Barometric Pressure
numeric (PB)
Range 400 to 800 mmHg
Resolution 1 mmHg
Peak Inspiratory Flow
numeric (PIF)
Range 0.3 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution 0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Peak Expiratory Flow
numeric (PEF)
Range 0.3 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution 0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Inspiration to
Expiration ratio numeric
(I:E)
Range 1: 0.2 to 9.9
Resolution 0.1
Spirometry Performance Specifications
C.O./CCO Performance Specifications
C.O. (right heart)
Range 0.1 to 20.0 l/min
Accuracy Instrument Specification (electrical): ±3% or 0.1 l/min
System Specification: ±5% or 0.2 l/min
Repeatability Instrument Specification (electrical): ±2% or 0.1 l/min
System Specification: ±3% or 0.1 l/min
C.O. (transpulmonary)
Range 0.1 to 25.0 l/min
Accuracy Instrument Specification (electrical): ±4% or 0.15 l/min
System Specification: ±5% or 0.2 l/min
Repeatability Instrument Specification (electrical): ±2% or 0.1 l/min
System Specification: ±3% or 0.1 l/min
EVLW Range 10 to 5000 ml
Standard Deviation 10% or 1 ml/kg
ITBV Range 50 to 6000 ml
Accuracy ±10% or 30 ml
Repeatability ±5% or 20 ml

Table of Contents