40 Specifications
475
Compatible Devices
M4605A Battery Specifications
The battery lifetime is 3 years from manufacturing date or 500 charge/discharge cycles.
Measurement Specifications
See the Appendix on Default Settings for a list of the settings the monitor is initially shipped with.
ECG/Arrhythmia/ST/QT
Complies with IEC 60601-2-25:1993 + A1:1999 /EN 60601-2-25:1995 + A1:1999,
IEC 60601-2-27:2005/EN 60601-2-27:2006, IEC 60601-2-51:2003 /EN 60601-2-51:2003 and AAMI
EC11/EC13:1991/2002.
Compatible Devices
Displays (must be approved for medical use)
M8031B XGA color 15" LCD touchscreen
M8033C SXGA color 17" LCD touchscreen
865299 SXGA color 19" LCD touchscreen
M4605A Battery Specifications
Physical Specifications
W x D x H 149 mm (5.866 in) x 89 mm (3.504 in) x 19.8 mm (0.78 in)
Weight 490 g (1.08 lb) per battery
Performance Specifications
Nominal Voltage 10.8 Volt
Rated Capacity at discharge C/5 6000 mAh
Continuous Discharge Capability 6.5 A
Environmental Specifications
Temperature Range Discharge 0 to 50°C (32 to 122°F)
Charge 0 to 50°C (32 to 122°F)
Storage and Transportation: -20 to 65°C (-4 to 140°F)
Humidity Range Operating: 15% to 95% Relative Humidity (RH)
Storage and Transportation: 5 % to 95 % Relative Humidity (RH)
Battery Type
Smart Battery 10.8 V, 6000 mAh, Lithium Ion
Safety
complies with UL 2054
Electromagnetic Compatibility (EMC)
complies with the requirements for FCC Type B computing Device, and
EN 61000-4-2 and EN 61000-3-2
Communication Standard
complies with the SMBus specification v 1.1