4 Patient Alarms and INOPs
91
Pressure INOPs
<SpO Label> Sensor Off
Numeric is replaced by -?-
INOP tone
The SpO
2
sensor is not properly applied to the patient. Apply the sensor following the instructions
supplied by the manufacturer.
<SpO Label>
Unkn.Sensor
Numeric is replaced by -?-
The connected sensor or adapter cable is not supported by the SpO
2
measurement. Use only specified
sensors and cables.
<SpO Label> Unplugged
Numeric is replaced by -?-
INOP tone
An SpO
2
measurement label has been deactivated, either by unplugging a module, or by deactivating
the label in the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module or reactivate the measurement label in
the Measurement Selection window.
<SpO Label> Upgrade
Numeric is replaced by -?-
Numeric is unavailable
The SpO
2
measurement is currently in upgrade mode. Monitoring is not possible in this mode.
INOP Message, Indication What to do
INOP Message, Indication What to do
ABP INOPs See <Press Label> INOPs
Ao INOPs See <Press Label> INOPs
ART INOPs See <Press Label> INOPs
BAP INOPs See <Press Label> INOPs
CPP Chk Sources
Numeric is replaced by -?-
INOP tone
Not all measurements or values required to perform the calculation are available. Check the
measurement sources.
CPP Chk Units
Numeric is replaced by -?-
The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
CPP Disabled CPP has been disabled, either in the setup in Configuration mode, or by loading settings with CPP
disabled. This can happen when changing from a pediatric or neonatal profile to an adult profile, as the
default profile settings have the CPP measurement disabled in the adult profile and enabled in the
pediatric and neonatal profiles. Enable CPP in the current profile to clear the INOP.
CVP INOPs See <Press Label> INOPs
FAP INOPs See <Press Label> INOPs
IC1 / IC2 INOPs See <Press Label> INOPs
ICP INOPs See <Press Label> INOPs
LAP INOPs See <Press Label> INOPs
P / P1 / P2 / P3 / P4 INOPs See <Press Label> INOPs
PAP INOPs See <Press Label> INOPs
PPV bad <Press Label>
Signal
The arterial pressure source selected for PPV is not providing a pulsatile signal.
PPV bad Signal
at Information Center
The arterial pressure source selected for PPV is not providing a pulsatile signal.
PPV Chk Sources The arterial pressure source selected for PPV is unplugged or switched off. When this INOP has
displayed for 1 minute PPV will be switched off.
<Press Label> Artifact
Numeric questionable
A non-physiological event (flush or blood sample) is detected. A resulting limit alarm or non-pulsatile
INOP will be suppressed.