41 Default Settings Appendix
514
NBP Default Settings
Temperature Default Settings
Low Alarm Delay (SAD) Short Short Short
Label SpO SpO SpO
Pulse settings
Pulse (SpO) OnOnOn
For all other Pulse default settings, refer to the Pulse Default Settings table.
SpO
2
Alarm Default Settings
Setting Adult Pedi Neo
NBP Settings
Factory defaults H10/H20/H30 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Mode Auto Auto Manual
Alarms from Sys. Sys. Sys.
High Limit 160/90 (110) 120/70 (90) 90/60 (70) 180/90 (110)
Low Limit 90/50 (60) 70/40 (50) 40/20 (24) 70/50 (65)
Alarms On On On
NBP OnOnOn
Repeat Time 10 min 10 min 10 min 3 min 3 min 3 min
Pulse (NBP) OnOnOn
Unit mmHg mmHg mmHg
Done Tone Off Off Off On On
Start Time Synchronize
d
Synchronize
d
Synchronize
d
NotSynchron
.
NotSynchron. NotSynchron
.
VP Pressure 60 mmHg 40 mmHg 30 mmHg
Reference Auscultatory Auscultatory Invasive
Aging Time 10 min 10 min 10 min
Color Red Red Red Magenta Magenta Magenta
Temp Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Low Limit 36 36 36 35 35 35
High Limit 39 39 39
Alarms On On On
Unit °C °C °C
Range 35...43 35...43 35...43
Color Green Green Green Light Green Light Green Light Green