41 Default Settings Appendix
519
tcGas Default Settings
Intravascular Oxygen Saturation Default Settings
tcGas Settings Factory Adult Factory Pedi Factory Neo
tcpO₂ High Limit 80 mmHg 80 mmHg 80 mmHg
tcpO₂ Low Limit 50 mmHg 50 mmHg 50 mmHg
tcpO₂ Alarms On On On
tcpCO₂ HighLimit 50 mmHg 50 mmHg 50 mmHg
tcpCO₂ Low Limit 30 mmHg 30 mmHg 30 mmHg
tcpCO₂ Alarms On On On
Site Timer 4.0 hours 4.0 hours 4.0 hours
Disable Timer Not Allowed Not Allowed Not Allowed
Heat Switch Off (i.e., after Site Timer elapsed) No No No
Transducer Temp. 43.0°C 43.0°C 43.0°C
CO₂ Correction (Severinghaus) On On On
MetabolismFactor 8 mmHg 8 mmHg 8 mmHg
tcGas Unit mmHg mmHg mmHg
Temperature Unit °C °C °C
tcpO₂ Color Blue Blue Blue
tcpCO₂ Color Green Green Green
SO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 70% 70% 70%
HR High Limit 80% 80% 80%
Alarms On On On
Light Intensity On On On
Color Yellow Yellow Yellow
Hb/Hct Entry Hct [%] Hct [%] Hct [%]
Factor Entry Disabled Disabled Disabled
SvO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 60% 60% 60%
High Limit 80% 80% 80%
Alarms On On On
Light Intensity On On On
Color Yellow Yellow Yellow
Hb/Hct Entry Hct [%]
Factor Entry Disabled