EasyManua.ls Logo

Philips MX400 - 41 Default Settings Appendix (cont. 3); tcGas Default Settings; Intravascular Oxygen Saturation Default Settings

Philips MX400
534 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
41 Default Settings Appendix
519
tcGas Default Settings
Intravascular Oxygen Saturation Default Settings
tcGas Settings Factory Adult Factory Pedi Factory Neo
tcpO High Limit 80 mmHg 80 mmHg 80 mmHg
tcpO Low Limit 50 mmHg 50 mmHg 50 mmHg
tcpO Alarms On On On
tcpCO HighLimit 50 mmHg 50 mmHg 50 mmHg
tcpCO Low Limit 30 mmHg 30 mmHg 30 mmHg
tcpCO Alarms On On On
Site Timer 4.0 hours 4.0 hours 4.0 hours
Disable Timer Not Allowed Not Allowed Not Allowed
Heat Switch Off (i.e., after Site Timer elapsed) No No No
Transducer Temp. 43.0°C 43.0°C 43.0°C
CO Correction (Severinghaus) On On On
MetabolismFactor 8 mmHg 8 mmHg 8 mmHg
tcGas Unit mmHg mmHg mmHg
Temperature Unit °C °C °C
tcpO Color Blue Blue Blue
tcpCO Color Green Green Green
SO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 70% 70% 70%
HR High Limit 80% 80% 80%
Alarms On On On
Light Intensity On On On
Color Yellow Yellow Yellow
Hb/Hct Entry Hct [%] Hct [%] Hct [%]
Factor Entry Disabled Disabled Disabled
SvO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 60% 60% 60%
High Limit 80% 80% 80%
Alarms On On On
Light Intensity On On On
Color Yellow Yellow Yellow
Hb/Hct Entry Hct [%]
Factor Entry Disabled

Table of Contents

Related product manuals