4 Patient Alarms and INOPs
103
Cableless Measurement Device INOPs
ΔSpO₂ Chk Units
Numeric is replaced by -?-
The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
Sp-vO₂ Chk Sources
Numeric is replaced by -?-
Not all measurements or values required to perform the calculation are available. Check measurement
sources.
Sp-vO₂ Chk Units
Numeric is replaced by -?-
The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
SVR Chk Sources
SVRI Chk Sources
Numeric is replaced by -?-
Not all measurements or values required to perform the calculation are available. Check measurement
sources.
SVR Chk Units
SVRI Chk Units
Numeric is replaced by -?-
The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
SVRSet CVP Used
SVRISet CVP Used
A CVP value is required for this calculation, but is not currently being measured. The monitor is using
the CVP value preset in the Setup SVR menu.
ΔTemp Chk Sources
Numeric is replaced by -?-
Not all measurements or values required to perform the calculation are available. Check measurement
sources.
ΔTemp Chk Units
Numeric is replaced by -?-
The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
INOP Message, Indication What to do
INOP Message, Indication What to do
cl NBP Batt Empty
!!cl NBP BattEmpty
!!!cl NBP BatEmpty
Corresponding INOP tone
The remaining battery time of the NBP Pod is below 30 minutes. Charge the battery.
cl NBP Batt Incomp
INOP tone
The battery in use with the NBP Pod is incompatible. Replace it with one approved for use with the
NBP Pod.
cl NBP Batt Low
INOP tone
The remaining battery time of the NBP Pod is below 2 hours.
cl NBP Batt Malf
INOP tone
There is a malfunction in the NBP Pod's battery system. Contact your service personnel.
cl NBP Batt Temp
INOP tone
The temperature of the battery in the NBP Pod is critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove the Pod from the patient and contact your
service personnel.
cl NBP Check Batt
INOP tone
The battery in the NBP Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
remain. Contact your service personnel to replace the battery.
cl NBP Chk Chrg IF The interface between the NBP Pod and the Charging Station is not working properly. Try cleaning the
contacts and reposition the Pod on the charging station.
cl NBP ChkSettings The NBP Pod's settings may have changed or become corrupt. Check the settings.
cl NBP Disconnect
INOP tone
The NBP Pod has lost the connection to the monitor.
cl NBP License Req The NBP Pod requires a valid software license for operating.
cl NBP No Cradle
INOP tone
The NBP Pod is not in its cradle.
cl NBP Remove
INOP tone
The temperature of the battery in the NBP Pod is too high. Remove the Cableless Measurement
Device from the patient and contact service personnel.