5 Managing Patients and Equipment
111
when no basic vitals (HR, RR, Pulse, SpO
2
, NBP) have been measured for a specified period
The pop-up window is entitled
Is this a New Patient?. The monitor offers a Yes key to stop
monitoring for the previous patient and begin monitoring a new patient and a
No key to continue
monitoring with the current patient data and settings.
The time periods for the three conditions can be configured independently.
Transferring Patients
To save you from having to enter the same patient data multiple times and enable patient transfer
without loss of data, information can be shared between Multi-Measurement Modules (MMS), patient
monitors, and Information Centers.
patient demographic information is shared between connected MMSs, patient monitors, and
Information Centers
measurement settings and calibration data can be uploaded from an MMS to a patient
monitor, if configured
trend information can be uploaded from an MMS to a patient monitor, if configured.
Different sets of patient and measurement-related data are stored in the monitor and the Multi-
Measurement Module. Understanding this will help you to understand what happens to patient data
when you transfer patients.
Patient Information Stored in Monitor Stored in X1 MMS and
extensions
Stored in X2 MMS and
extensions
Patient demographics (name,
DOB, patient IDs)
yes yes yes
Monitor settings (alarm
pause time, alarm volume)
yes no no
Measurement settings for all
measurements (alarm limits,
measurement on/off, etc.)
yes yes, for all MMS and MMS
extension measurements
yes, for all MMS and
MMS extension
measurements
Trend data
yes, for all MMS, MMS
extension and measurement
module measurements (up to
a maximum of 24 or 32 or 50
or 100, depending on your
database configuration)
yes, most recent 8 hours of
information, for all MMS
and extensions
measurements
yes, most recent 8 hours
of information. A
standalone X2 stores all
MMS and MMS extension
measurements (up to a
maximum of 50
measurements).
During Companion Mode
the X2 stores all MMS and
MMS extension
measurements and
measurements collected
by the host monitor from
measurement modules (up
to a maximum of 50
measurements)