41 Default Settings Appendix
517
Cardiac Output Default Settings
Sys. Low Clamp/Dia. Low Clamp (Mean
Low Clamp)
0/-5 (-5) 0/-5 (-5) 0/-5 (-5)
Scale 30 30 30
Mean only Yes Yes Yes
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Magenta Magenta Magenta
ICP, IC1, IC2 Settings
Factory Defaults
Adult Pedi Neo
C.O. Settings Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Auto-Calibration On On On
Tblood HighLimit 39.0°C 39.0°C 39.0°C
Tblood Low Limit 36.0°C 36.0°C 36.0°C
Tblood Alarms On On On
Temperature Unit °C °C °C
Color Green Green Green Yellow Yellow Yellow
CCO/CCI Settings Factory Adult Factory Pedi Factory Neo
Settings common to CCO and CCI
Alarms from CCO CCO CCO
CCO from ABP ABP ABP
Color Green Green Green
CCO
Alarms On On On
CCO High Limit 8.5 l/min 3.7 l/min 1.3 l/min
CCO Low Limit 4.0 l/min 2.6 l/min 0.3 l/min
CCI
Alarms On On On
CCI High Limit 4.3 l/min/m
2
4.3 l/min/m
2
5.2 l/min/m
2
CCI Low Limit 2.0 l/min/m
2
2.6 l/min/m
2
1.2 l/min/m
2