20 Monitoring NMT
298
Changing the NMT Measurement Settings
Selecting the NMT Stimulation Mode
In the Setup NMT window:
1 Select Stimulation to enter the Stimulation window.
2 Choose the Stimulation mode required:
–
TOF
– PTC
– Twitch (if configured)
–
DBS (if configured)
Measurement Mode
Allows the choice between manual and automatic NMT measurements.
In the
Setup NMT window:
1 Select Mode to open the Mode window.
2 Select Auto or Manual mode.
NOTE
Auto mode is not available if the PTC or DBS stimulation mode is selected.
Selecting the Repeat Time (only available in Auto mode)
Repeat Time allows you to select the time intervals between automatic NMT measurements.
NOTE
This function is not available in the PTC and DBS mode.
In the
Setup NMT window:
1 Select Repeat Time to open the Repeat Time window.
2 Select the required setting.
Selecting the Stimulation Current
To set the stimulation current.
In the
Setup NMT window:
1 Select Current to open the Current window.
2 Select the required setting.
3 If you are modifying a previously calibrated current, select Confirm.
CAUTION
Changing the stimulation current after calibration invalidates the stored reference data.