22
307
22Using a Telemetry Device and a
Monitor (PIIC only)
It is possible to assign a monitor and a telemetry device to the same patient, resulting in the
information from both devices being combined in one sector at the Information Center. This is called
"pairing" and when a telemetry device and a monitor are paired, the measurement data from the
telemetry device will be displayed on the monitor screen.
A telemetry device can be:
any Philips telemetry device (only for indirect connections, see below).
a TRx/TRx+ Intellivue Telemetry system transceiver.
an X2 Multi measurement module with a telemetry interface, declared as a telemetry device and
with a telemetry equipment label.
For complete user information, including accessory listings and specifications, for the M4841A/
M4851A TRx/TRx+ IntelliVue Telemetry Transceiver, refer to the Instructions for Use provided with
the device.
How Can You Combine Devices?
•With an indirect connection, using standard telemetry transmission - the data are sent to the
monitor via the Information Center and arrive with a delay of several seconds at the monitor.
•With a direct connection to the monitor
an X2 is connected via MSL cable to the monitor, or
a telemetry transceiver with a short range radio adapter is assigned to an X2 which is
connected (Companion Mode) to an MX400/450, MX500/550, or MX600/700/800 host
monitor
the data arrive with a minimal delay on the monitor screen.
Indirect Connection - Manual Pairing
The telemetry device can be paired with the monitor at the Information Center or at the monitor. For
detailed information regarding pairing and configuration at the Information Center, see the
Information Center Instructions for Use.
To pair the monitor with a telemetry device at the monitor:
1 Select Main Setup then Measurements