23 Trends
316
To switch off automatic optimum scaling,
•in the
Segment menu, select Optimum Scale again.
Trend Group
To add or remove measurements for this trend group or change the order of the existing
measurements:
1 In the Segment menu, select Change Group.
2 Use the Add, Change, Sort Up and Sort Down pop-up keys to change or re-order the group as
required.
No. of Segments
In an embedded graphical trend window, you can select the number of segments to be displayed in the
Segment menu:
•In the
Segment menu, select No. of Segments.
Trend Groups
The measurements grouped in trend groups define the trends displayed together in the Vital Signs or
Graphical Trends windows and printed in trends reports and recordings. The sequence of the
measurements in the group defines the order in which they are displayed. Measurements listed
between dotted line separators are displayed overlapping. The trend group
All contains all available
measurements, you cannot change the order or selection of this group.
To change the selection of measurements in a trend group, either use the
Change Group setting in the
Segment menu or:
1 Select Main Setup, Trends, then Trend Groups.
2 Select the Trend Group you want to change and use the pop-up keys to Add, Change, or Delete
the selection of measurements trended.
To temporarily change the order in which the measurements are displayed in a group,
1 Select Main Setup, Trends, then Trend Groups.
2 Select the Trend Group and then the measurement you want to move and use the Sort Up/Sort
Down
pop-up keys.
Trend Interval
The trend interval defines the resolution of trend data shown on the Screen. High-resolution data is
especially suited for neonatal applications, where the clinical situation may change very quickly. In
adult monitoring, where the patient's status typically changes more gradually, a longer trend may be
more informative.
To set the trend resolution, in the
Vital Signs or Graphical Trends window,
• Select the
Select Interval pop-up key and then select the required interval from the list.
Trend Priority
The monitor stores trend information for all monitored measurements, if configured to do so. (Data
from VueLink or IntelliBridge modules cannot be included in trends when the label is a free-text label.)