40 Specifications
479
SpO2
Unless otherwise specified, this information is valid for SpO
2
measured using the M3001A and
M3002A Multi-measurement modules and the M1020B measurement module. The SpO
2
Performance
Specifications in this section apply to devices with Philips SpO
2
technology. For SpO
2
Performance
Specifications valid for other SpO
2
technologies, refer to the instructions for use provided with these
devices.
Complies with ISO 9919:2005 / EN ISO 9919:2009 (except alarm system; alarm system complies with
IEC 60601-2-49:2001).
Measurement Validation: The SpO
2
accuracy has been validated in human studies against arterial
blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically
distributed, only about two-thirds of the measurements can be expected to fall within the specified
accuracy compared to CO-oximeter measurements.
Display Update Period: Typical: 2 seconds, Maximum: 30 seconds. Maximum with NBP INOP
suppression on: 60 seconds.