40 Specifications
480
SpO
2
Performance Specifications
SpO2
The specified accuracy is
the root-mean-square
(RMS) difference between
the measured values and
the reference values
Range 0 to 100%
Option #A01
Accuracy
Philips Reusable Sensors:
M1191A, M1191AL, M1191B, M1191BL, M1192A = 2 % (70 % to 100 %)
M1193A, M1194A, M1195A, M1196A, M1196S = 3 % (70 % to 100 %)
M1191T, M1192T, M1193T (Adult), M1196T = 3% (70% to 100%)
M1193T (Neonate) = 4 % (70% to 100%)
Philips Disposable Sensors with M1943A(L):
M1132A, M1133A (adult/infant), M1134A (adult/infant) = 2 %
M1901B, M1902B, M1903B, M1904B, M1131A, M1133A (neonate), M1134A
(neonate) = 3 % (70% to 100%)
NellcorPB
®
Sensors with M1943A(L):
MAXA, MAXAL, MAXP, MAXI, MAXN, D-25, D-20, I-20, N-25, Oxicliq A,
P, I, N = 3 % (70% to 100%)
Masimo Reusable Sensors
®
with LNOP MP12 or LNC MP10:
LNOP DC-I, LNOP DC-IP, LNOP YI (adult/pedi/infant), LNCS DC-1,
LNCS DC-IP, LNCS YI (adult/pedi/infant) = 2% (70% to 100%)
LNOP YI (neonate), LNCS YI (neonate) = 3% (70% to 100%)
LNOP TC-I, LNCS TC-I: 3.5 % (70% to 100%)
Masimo Disposable Sensors
®
with LNOP MP12 or LNC MP10:
LNOP Adt, LNOP Adtx, LNOP Pdt, LNOP Pdtx, LNOP Inf-L, LNOP Neo-L
(adult), LNCS Adtx, LNCS Adtx-3, LNCS Pdtx, LNCS Pdtx-3, LNCS Inf,
LNCS Inf-3, LNCS Neo (adult), LNCS Neo-3 (adult) = 2% (70% to 100%)
LNOP Neo-L (neonate), LNOP NeoPt-L, LNCS Neo (neonate), LNCS Neo-3
(neonate), LNCS NeoPt, LNCS NeoPt-3 = 3% (70% to 100%)
Option #A02
Accuracy
Philips Reusable Sensors
M1193AN = 3 % (70% to 100%)
Philips Disposable Sensors with M1943NL:
M1901B (Adult), M1902B, M1903B, M1904B = 2% (70% to 100%)
M1901B (Neonate) = 3%
Nellcor Sensors with M1943NL:
MAXFAST, MAXA, MAXAL, MAXP, MAXI, MAXN (Adult) = 2 % (70 % to
100 %)
SC-A, Oxicliq A, P, I, N (Adult) = 2.5% (70% to 100%)
MAXN (Neonate), Dura-Y D-YS (Infant to Adult), DS-100A-1, Oxiband OXI-
A/N (Adult), OXI-P/I = 3 % (70% to 100%)
MAXR, Oxicliq N (Neonate), SCNEO-I (Neonate), SC-PR-I (Neonate), Dura-
Y D-YS with D-YSE or D-YSPD clip = 3.5 % (70% to 100%)
Dura-Y D-YS (Neonate), Oxiband OXI-A/N (Neonate) = 4% (70% to 100%)
Resolution 1%
Averaging Time 5, 10 or 20 seconds (user adjustable)
Pulse Range 30 to 300 bpm
Accuracy ±2% or 1 bpm, whichever is greater
Resolution 1 bpm
Sensors Wavelength range: 500 to 1000 nm
Emitted Light Energy: ≤15 mW
Information about the wavelength range can be especially useful to clinicians
(for instance, when photodynamic therapy is performed)