5 Managing Patients and Equipment
117
PIIC
• When a monitor is connected to an Information Center by the wireless IntelliVue Instrument
Telemetry interface, the patient data will automatically be merged in the case of a transfer. This
means there is no patient discharge at the monitor and settings and trend data will be retained. You
will see a message on the monitor and the
Patient Demographics window will automatically
appear so that you can check the data and change it if necessary.
In the case where an X2 with an IntelliVue Instrument Telemetry interface is declared as a "telemetry
device" at the Information Center and is connected to a host monitor, it is important to resolve an
existing mismatch between the monitor and the Information Center before disconnecting the X2.
Failure to do so discharges the X2 and synchronizes the demographics and settings to the
Information Center.
Manually Resolving Patient Mismatch
The patient mismatch is indicated by question marks (???) beside the questionable fields in the Monitor
Info Line and in the
Select Patient window. The monitor displays a message such as Different
patients in Central and Monitor
.
The
Select Patient window automatically opens so you can decide which patient data to use. You do
not have to resolve the mismatch immediately, but the indicators remain until you do. After 5 minutes
a
Check Patient ID INOP will appear to remind you to resolve the mismatch.
For some common mismatch situations, the monitor will simplify the resolution by suggesting a
solution for the mismatch. For example, when a patient arrives after transport and the
Transfer key
has been selected, the monitor will show this patient's data and ask
Complete transfer of this
patient?
. You can then select Yes to complete the transfer. If you select No you will go to the Select
Patient
window.
There can be up to three sets of demographic data in the
Select Patient window if the patient is
different in the Information Center, monitor, and MMS.
After you resolve the mismatch, the monitor displays a confirmation window that shows the patient
that has been selected and where data will be erased, if applicable. Confirm your choice. The monitor
automatically displays the
Patient Demographics window after confirmation. Verify that the settings
shown are correct for the patient.
Gender, date of birth, height, weight, and nursing notes do not generate a mismatch. If these fields are
different on different devices, the monitor resolves them itself. For example, it may take date of birth
from the Information Center, whilst taking gender from the MMS. Always check the Patient
Demographics after combining patients, to ensure that you are satisfied with the results. Change them
if necessary.
WARNING
After resolving a patient mismatch, check that the monitor settings (especially patient category, paced
status and alarm limits) are correct for the patient.
Patient Mismatch - If One Set of Patient Data is Correct
If there is a mismatch between an Information Center and a monitor, or a monitor and an MMS,
choose the data set you want to continue using for this patient by selecting one of the sectors in the
Select Patient window.