41 Default Settings Appendix
511
ST and QT Default Settings
SVT On On On
Afib On On On
Irregular HR On On On Off Off Off
HR Alarms Short Yellow Short
Yellow
Short
Yellow
TimeOut 1st 3 min 3 min 3 min
TimeOut 2nd 10 min 10 min 10 min
ArrhyOff Message Yes Yes Yes
Some ECG Al INOP On On On Off Off Off
Arrhythmia Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Lead-independent ST Settings
Factory Defaults
Factory Adult Factory Pedi Factory Neo
ISO/J Point detection Auto Auto Auto
Alarms (ST) On On On
STE Off Off Off
Alarms (STE) On On On
ST Analysis On Off Off
ST-Index Off Off Off
ISO Point -80 ms -80 ms -80 ms
J Point 48 ms 48 ms 48 ms
ST Offset J+60 J+60 J+60
ST Settings - Lead I, II, III, V, aVR, aVL, V1-
6, MCL
Factory Defaults
Factory Adult Factory Pedi Factory Neo
ST (Label) High Alarm Limit +1.0 mm +1.0 mm +1.0 mm
ST (Label) Low Alarm Limit -1.0 mm -1.0 mm -1.0 mm
ST Settings - Lead V, MCL
ST (Label) High Alarm Limit +2.0 mm +2.0 mm +2.0 mm
ST (Label) Low Alarm Limit -2.0 mm -2.0 mm -2.0 mm
STE Settings - Limb leads
STE Male Limb, Alarm Limit 1.0 mm 1.0 mm 1.0 mm
STE Female Limb, Alarm Limit 1.0 mm 1.0 mm 1.0 mm
ST Settings - lead V2, V3
STE Male V2,V3, Alarm Limit
STE Female V2,V3, Alarm Limit
STE Settings - V1, V4-V6
STE Male V1,V4-6, Alarm Limit 1.0 mm 1.0 mm 1.0 mm