40 Specifications
467
Trends Resolution 12, 16, 24 or 32 numerics @ 12 sec, 1 minute, 5 minute
resolution
Information Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
neonatal extended 12 numerics, 24 hours @ 12 secs or
32 numerics 32 hours @ 1 minute
intensive care extended: 16 numerics 120 hours @ 5 minutes
anesthesia extended 32 numerics 9 hours @ 12 seconds
High Res Trend Waves Measurements available HR, SpO
2
, Resp, tcpO
2
, Pulse, Perf, tcpCO
2
, CO
2
, ABP, PAP,
CVP, ICP, CPP, BIS, CCO, AWP, Anesthetic Agents, Delta
SpO
2
, inO
2
Resolution Measurement samples are taken at a resolution of four samples
per second
Update speed waves are drawn at a speed of 3 cm/minute
Events Information trigger condition and time, event classification and associated
detailed view of episode data
Episode data configurable, either:
4 minutes of high resolution trend or
20 minutes of numerics trend @ 12 sec. resolution or
15 seconds of 4 waves @ 125 samples/sec. (Snapshot)
including all current numerics, alarms and inops
Capacity (max.) 25 or 50 events for either 8 or 24 hours
Alarm Signal System alarm delay
The system alarm delay is the
processing time the system needs
for any alarm to be indicated on
the monitor, after the
measurement has triggered the
alarm.
less than 4 seconds
Delay for alarm availability on the
network
This is the time needed after alarm
indication on the monitor until the
alarm signal is available on the
network, to the IntelliVue
Information Center or for
transmission to other systems.
less than 5 seconds
Pause duration 1,2,3 minutes or infinite, depending on configuration
Extended alarm pause 5 or 10 minutes
Sound pressure range min. 0 dB(A)
max. 45-85 dB(A)
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity 300 items
Real Time Clock Range from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy better than 4 seconds per day
Hold Time infinite if powered by AC; otherwise at least 24 hours
MX600/700/800