8 Monitoring Respiration Rate (Resp)
189
This is because of the higher internal impedance of the OR cable set, required for use if electrosurgery
is being performed.
Rate adaptive pacemakers
Implanted pacemakers which can adapt to the Minute Ventilation rate may occasionally react on the
Impedance measurement used by patient monitors for the determination of the Resp value and
execute pacing with the maximum programmed rate. Switching off the Resp measurement can prevent
this.