System Configuration
R&S
®
ESR
511User Manual 1175.7068.02 ─ 12
1 = Name and comment of the transducer set
2 = Unit of the transducer set
3 = Button to turn the transducer break on and off
4 = Table of transducer set ranges and corresponding frequencies and transducer factors
5 = Start and stop frequency of the currently selected range
6 = List of transducer ranges
For more information on using and designing transducer sets, see "Transducer sets"
on page 175.
Opening the dialog box also opens an additional softkey menu that contains functional-
ity that you need to design a transducer set.
This softkey menu provides the function to
Insert ranges up to a maximum of 10 ("Insert Range" softkey)
Delete ranges that already exist ("Delete Range" softkey)
Add up to eight transducer factors to a range ("Add Factor" softkey)
Replace a transducer factor that has been assigned with another one ("Change
Factor)" softkey)
Remove a transducer factor from the transducer set ("Remove Factor" softkey)
Save the contents of the transducer set ("Save Set" softkey)
Remote command:
[SENSe:]CORRection:TSET:SELect on page 956
[SENSe:]CORRection:TSET:COMMent on page 955
[SENSe:]CORRection:TSET:UNIT on page 956
[SENSe:]CORRection:TSET:BREak on page 954
[SENSe:]CORRection:TSET:RANGe<range> on page 955
[SENSe:]CORRection:TSET:DELete on page 955
Alignment
Opens a submenu with the available functions for recording, displaying and activating
the data for self-alignment.
The R&S ESR determines the correction data and characteristics required for the
alignment automatically. It compares the results at different settings with the known
characteristics of the high-precision calibration signal source at 65.83 MHz. The cor-
rection data is stored as a file on the instrument. You can display it using the "Show
Align Results" on page 512 softkey.
The submenu contains the following commands:
"Self Alignment" on page 511
"Show Align Results" on page 512
"Touch Screen Alignment" on page 512
Self Alignment ← Alignment
Starts the recording of correction data of the instrument. If the correction data acquisi-
tion has failed or if the correction values are deactivated, a corresponding message is
displayed in the status field.
As long as the self-alignment data is being collected, you can cancel the procedure
using the "Abort" button.
Instrument Setup and Interface Configuration – SETUP Key