System Configuration
R&S
®
ESR
514User Manual 1175.7068.02 ─ 12
LXI functionality is available only for user accounts with administrator rights.
Remote command:
SYSTem:LXI:PASSword on page 990
Description ← LXI ← General Setup
Opens a dialog box to view or change the LXI instrument description. This description
is used on some of the LXI web sites.
By default, the description is "Signal Analyzer".
LXI functionality is available only for user accounts with administrator rights.
Remote command:
SYSTem:LXI:MDEScription on page 990
LAN Reset ← LXI ← General Setup
Resets the LAN configuration to a state required by the LXI standard. For example, the
TCP/IP mode is set to "DHCP" and "Dynamic DNS" and "ICMP Ping" are enabled. In
addition, the R&S ESR sets the password and the instrument description to their initial
states (see "Password" on page 513 and "Description" on page 514 softkeys).
Only user accounts with administrator rights can reset the LAN configuration.
Remote command:
SYSTem:LXI:LANReset on page 990
GPIB ← General Setup
Opens a submenu to set the parameters of the remote control interface.
GPIB Address ← GPIB ← General Setup
Opens an edit dialog box to enter the GPIB address. Values from 0 to 30 are allowed.
The default address is 20.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess on page 989
ID String Factory ← GPIB ← General Setup
Selects the default response to the *IDN? query.
Remote command:
SYSTem:IDENtify:FACTory on page 926
ID String User ← GPIB ← General Setup
Opens an edit dialog box to enter a user-defined response to the *IDN? query. Max.
36 characters are allowed.
Remote command:
SYSTem:IDENtify[:STRing] on page 926
Compatibility Mode ← GPIB ← General Setup
Sets the R&S ESR in a state compatible to previous R&S devices, enabling the usage
of existing external control applications. In particular, the number of measurement
points and available bandwidths are adjusted to the settings of other devices.
Instrument Setup and Interface Configuration – SETUP Key

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals