Remote Control
R&S
®
ESR
853User Manual 1175.7068.02 ─ 12
Manual operation: See "Lin" on page 393
See "Log" on page 393
See "Power" on page 394
[SENSe:][WINDow:]DETector<trace>[:FUNCtion] <Function>
This command selects the detector for the data acquisition in the selected trace.
Suffix:
<trace>
.
1...4
Selects the trace.
<trace> 1...6
trace
Parameters:
<Function> APEak | NEGative | POSitive | SAMPle | RMS | AVERage |
QPEak | CAVerage | CRMS
*RST: APEak
Example:
DET POS
Sets the detector to "positive peak".
Manual operation: See "Trace 1 - 6" on page 199
See "Auto Peak" on page 392
See "Positive Peak" on page 392
See "Negative Peak" on page 392
See "Sample" on page 392
See "RMS" on page 392
See "Average" on page 392
See "Quasipeak" on page 392
See "CISPR Average" on page 392
See "RMS Average" on page 393
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]:AUTO <State>
This command either couples the detector to the current trace setting or turns coupling
off.
Suffix:
<trace>
.
1...6
trace
Parameters:
<State> ON | OFF
*RST: ON
Example:
DET:AUTO OFF
Manual operation: See "Auto Select" on page 391
Remote Commands in Spectrum Analyzer Mode