6-14 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
d. Carefully pull the reagent manifold out slightly, then straight down, to avoid the
sample connector (item 3, Figure 6-1).
Figure 6-1 Reagent Manifold and Tubing Connections
5. Place the reagent manifold on a stable, flat surface.
6. Fill the plastic syringe with water and place it against the sensor waste port (item 4,
Figure 6-2).
7. Press the syringe plunger gently to insert water into the CO-ox waste port, while
holding gauze against the end of the tubing that is connected to the
CO-ox waste exit port (item 2, Figure 6-2), to catch the obstruction and any excess
liquid.
1 Fitting to which the reagent manifold tubing and CO-ox pump tubing are connected
(Rapidlab 1245 and 1265 systems)
2 Reagent manifold tubing that connects the fitting to the port on the top, left-side of the
reagent manifold (Rapidlab 1245 and 1265 systems)
3 Sample connector
4 Waste tubing connected to port on top, right-side of the reagent manifold
5 Two latches that secure the reagent manifold to the system
BIOHAZARD: Discard the gauze according to the established biohazard disposal
practices of your workplace.