6-80 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
No AQC
Cartridge
The system does not detect the
AutomaticQC cartridge after
replacing the cartridge or at
power up.
If the AutomaticQC cartridge
was just installed, the cartridge
may not be installed correctly or
it could be a system problem.
Replace the AutomaticQC
cartridge as described Replacing
the AutomaticQC Cartridge
page 2-3.
If you are not able to replace the
cartridge, continue running
patient samples with the
AutomaticQC turned off. For
information about AutomaticQC
Options, refer to Setting up
AutomaticQC Schedule
page 8-5.
No Paper in
Printer
The printer is out of paper. Install a new roll of paper.
Refer to Replacing the Printer
Paper‚ page 5-42.
If necessary, use the Recall menu
to locate and print results or data
that were not printed while the
printer was out of paper.
No Printer
Configured
You selected the external printer
for a printout but you have not
configured Microsoft Windows
with a default external printer.
Configure Windows with a
default printer.
No R Cartridge The system does not detect the
reagent cartridge.
The cartridge may not be
installed correctly.
Reinstall the cartridge.
Replace the reagent cartridge.
Refer to Replacing the Reagent
Cartridge‚ page 2-3.
No Sample
Detected
The system is unable to detect a
sample at FD1.
This may be caused if you do not
insert a sample device before
starting sample analysis, by air
leaks in tubing, or if tubing is
disconnected.
Check tubing connections and
check for leaks in the sample
tubing.
No Waste
Bottle
Detected
The system does not detect the
waste bottle.
Place the waste bottle on the
system.
Refer to Emptying the Waste
Bottle‚ page 2-2
Message Probable Cause Corrective Action