Rapidlab 1200 Operator’s Guide: Troubleshooting 6-83
02087462 Rev. V
User Canceled
Calibration
The operator interrupted a
calibration to perform another
action, for example to load a
STAT test.
No action necessary.
W Cartridge
Expired
The wash cartridge has exceeded
its operating life or is depleted.
Replace the cartridge. Refer to
Replacing the Wash Cartridge
page 2-2.
W Cartridge
Not Valid
An expired wash cartridge was
installed or the cartridge was
installed after being removed for
more than 6 hours.
The system is unable to use the
cartridge.
Replace the cartridge. Refer to
Replacing the Wash Cartridge
page 2-2.
Waste Full The waste bottle is full and can
no longer accept waste.
Replace the waste bottle. Refer to
Emptying the Waste Bottle
page 2-2.
Waste is
Almost Full
The waste bottle is
approximately 70% full.
Replace the waste bottle. Refer to
Emptying the Waste Bottle
page 2-2.
You can replace the waste bottle
now or wait until the system
prompts you.
Message Probable Cause Corrective Action