Rapidlab 1200 Operator’s Guide: System Specifications E-15
02087462 Rev.V
2 Syringe 48 0.39 22.7 21.5 105.7 1.7
3 Syringe 45 0.59 26.7 28.6 93.3 2.2
4 Syringe 48 0.81 36.3 35.4 102.5 2.2
5 Syringe 48 1.08 50.8 50.1 101.4 2.1
6 Syringe 45 1.79 70.7 71.5 98.8 2.5
7 Syringe 48 1.97 99.8 99.1 100.7 2
8 Syringe 48 4.35 130.3 135.9 95.9 3.3
9 Syringe 47 5.59 222.7 226.5 98.3 2.5
1Micro-
capillary
12 0.42 7.2 7.1 100.7 5.8
2Micro-
capillary
12 0.74 21.3 21.4 99.5 3.5
3Micro-
capillary
12 0.42 24.5 28.5 85.9 1.7
4Micro-
capillary
12 0.33 33.3 35.7 93.4 1
5Micro-
capillary
12 0.85 49.5 49.9 99 1.7
6Micro-
capillary
12 1.12 67.6 71.3 94.9 1.7
7Micro-
capillary
12 1.23 100.1 99.9 100.2 1.2
8Micro-
capillary
12 2.22 134.2 135.6 99 1.7
9Micro-
capillary
12 6.69 228.9 228.3 100.2 2.9
1Micro-
syringe
24 0.36 6.9 7.2 96.3 5.3
2Micro-
syringe
24 0.38 21.4 21.2 100.7 1.8
3Micro-
syringe
24 1.26 26.2 28.3 92.7 4.8
4Micro-
syringe
24 0.82 34.1 35.8 95.2 2.4
5Micro-
syringe
24 1.09 50.5 50.1 100.8 2.2
6Micro-
syringe
24 2.02 69.8 70.8 98.7 2.9
Table E-12 Rapidlab 1240 Recovery and Precision Testing—pCO
2
Level Mode n WRSD
*
Observed Expected Recovery WRCV