E-16 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
7Micro-
syringe
24 1.91 101.7 100.2 101.5 1.9
8Micro-
syringe
24 2.97 138.3 136 101.7 2.1
9Micro-
syringe
24 6.74 241.4 229 105.4 2.8
* WRSD = within-run standard deviation
Table E-13 Rapidlab 1240 Recovery and Precision Testing—pO
2
Level Mode n WRSD
*
Observed Expected Recovery WRCV
1 Capillary 24 0.34 27.9 28.6 97.7 1.2
2 Capillary 24 0.35 52 50 104 0.7
3 Capillary 24 0.89 87.7 85.5 102.5 1
4 Capillary 24 1.47 101.4 107.1 94.7 1.5
5 Capillary 24 0.79 153.7 149.6 102.7 0.5
6 Capillary 24 4.02 198 199.9 99 2
7 Capillary 24 1.53 254.1 249.4 101.9 0.6
8 Capillary 24 4.87 374.8 378.4 99 1.3
9 Capillary 24 10.28 630 649.7 97 1.6
1 Syringe 45 0.36 28.1 28.6 98 1.3
2 Syringe 48 0.42 52.5 50.1 104.9 0.8
3 Syringe 48 0.6 88.2 84.9 103.9 0.7
4 Syringe 48 1.15 103.2 107.3 96.2 1.1
5 Syringe 48 1.13 153.7 148.6 103.4 0.7
6 Syringe 45 2.26 196.2 200.3 98 1.2
7 Syringe 47 3.81 253.9 247.7 102.5 1.5
8 Syringe 48 4.69 379.5 379.2 100.1 1.2
9 Syringe 48 9.97 636.6 651.1 97.8 1.6
1Micro-
capillary
12 0.21 30.7 28.5 107.5 0.7
2Micro-
capillary
12 0.35 54 49.9 108 0.6
3Micro-
capillary
12 0.7 95.1 85.6 111 0.7
4Micro-
capillary
12 0.64 106.7 107 99.7 0.6
Table E-12 Rapidlab 1240 Recovery and Precision Testing—pCO
2
Level Mode n WRSD
*
Observed Expected Recovery WRCV