EasyManua.ls Logo

Siemens Rapidlab 1200 - Page 382

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-22 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
1 pH Only
Capillary
24 0.007 7.024 7.032 99.88 0.10
2 pH Only
Capillary
24 0.004 7.198 7.202 99.94 0.05
3 pH Only
Capillary
24 0.008 7.362 7.354 100.11 0.10
4 pH Only
Capillary
24 0.003 7.414 7.410 100.05 0.04
5 pH Only
Capillary
24 0.005 7.542 7.534 100.11 0.07
6 pH Only
Capillary
24 0.004 7.653 7.630 100.30 0.05
1 pH Only
Syringe
48 0.004 7.016 7.032 99.77 0.06
2 pH Only
Syringe
48 0.003 7.189 7.202 99.82 0.05
3 pH Only
Syringe
48 0.002 7.350 7.354 99.95 0.03
4 pH Only
Syringe
48 0.003 7.389 7.410 99.72 0.04
5 pH Only
Syringe
48 0.002 7.517 7.534 99.78 0.03
6 pH Only
Syringe
48 0.003 7.631 7.630 100.01 0.04
* WRSD = within-run standard deviation
Table E-19 Rapidlab 1245 Recovery and Precision Testing—pCO
2
Level Mode n WRSD
*
Observed Expected Recovery WRCV
1 Capillary 24 0.16 7.5 7.1 104.7 2.1
2 Capillary 24 0.29 22.4 21.4 104.5 1.3
3 Capillary 24 0.56 27.3 28.6 95.4 2.1
4 Capillary 24 0.63 35.3 35.6 99.2 1.8
5 Capillary 24 0.94 49.6 50.0 99.2 1.9
6 Capillary 24 1.39 69.5 71.4 97.4 2.0
7 Capillary 23 3.58 95.4 99.7 95.6 3.8
8 Capillary 24 2.98 133.3 135.7 98.3 2.2
9 Capillary 24 3.27 215.7 228.0 94.6 1.5
1 Syringe 48 0.22 7.4 7.2 104.0 3.0
Table E-18 Rapidlab 1245 Recovery and Precision Testing—pH
Level Mode n WRSD
*
Observed Expected Recovery WRCV

Table of Contents