Loading...
Rapidlab 1200 Operator’s Guide: System Specifications E-23
02087462 Rev.V
2 Syringe 48 0.32 22.3 21.5 104.0 1.4
3 Syringe 48 0.47 27.4 28.6 95.6 1.7
4 Syringe 48 1.26 35.5 35.4 100.2 3.6
5 Syringe 48 1.47 50.9 50.1 101.6 2.9
6 Syringe 48 1.60 69.4 71.5 97.0 2.3
7 Syringe 48 3.69 98.7 99.1 99.7 3.7
8 Syringe 48 7.85 130.3 135.9 95.9 6.0
9 Syringe 45 4.22 220.7 226.4 97.5 1.9
1Micro-
capillary
12 0.30 7.4 7.1 103.5 4.1
2Micro-
capillary
12 0.37 22.8 21.4 106.4 1.6
3Micro-
capillary
12 0.58 27.2 28.5 95.3 2.1
4Micro-
capillary
12 0.52 36.7 35.7 103.0 1.4
5Micro-
capillary
12 1.25 52.3 49.9 104.6 2.4
6Micro-
capillary
12 2.41 69.3 71.3 97.2 3.5
7Micro-
capillary
12 2.35 102.9 99.9 103.0 2.3
8Micro-
capillary
12 3.29 135.8 135.6 100.1 2.4
9Micro-
capillary
12 6.71 222.4 228.3 97.4 3.0
1Micro-
syringe
24 0.26 7.2 7.2 101.2 3.6
2Micro-
syringe
24 0.41 21.1 21.2 99.2 2.0
3Micro-
syringe
24 0.62 25.6 28.3 90.5 2.4
4Micro-
syringe
24 0.63 34.5 35.8 96.5 1.8
5Micro-
syringe
24 0.92 49.1 50.1 97.9 1.9
6Micro-
syringe
24 1.88 64.6 70.8 91.2 2.9
Table E-19 Rapidlab 1245 Recovery and Precision Testing—pCO
2
Level Mode n WRSD
*
Observed Expected Recovery WRCV

Table of Contents