E-24 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
7Micro-
syringe
24 1.80 98.0 100.2 97.8 1.8
8Micro-
syringe
24 2.55 127.8 136.0 94.0 2.0
9Micro-
syringe
24 5.59 224.7 229.0 98.1 2.5
* WRSD = within-run standard deviation
Table E-20 Rapidlab 1245 Recovery and Precision Testing—pO
2
Level Mode n WRSD
*
Observed Expected Recovery WRCV
1 Capillary 24 0.26 28.8 28.6 100.9 0.9
2 Capillary 24 0.37 51.7 50.0 103.5 0.7
3 Capillary 24 1.55 88.9 85.5 104.0 1.7
4 Capillary 24 1.32 103.5 107.1 96.7 1.3
5 Capillary 23 1.33 152.2 149.6 101.8 0.9
6 Capillary 24 2.49 199.0 199.9 99.5 1.2
7 Capillary 24 10.54 259.4 249.4 104.0 4.1
8 Capillary 24 3.90 378.5 378.4 100.0 1.0
9 Capillary 24 4.60 656.6 649.7 101.1 0.7
1 Syringe 48 0.29 27.6 28.6 96.5 1.0
2 Syringe 48 0.39 51.5 50.1 102.9 0.8
3 Syringe 48 0.88 87.1 84.9 102.5 1.0
4 Syringe 48 0.94 104.8 107.3 97.7 0.9
5 Syringe 48 0.97 153.0 148.6 103.0 0.6
6 Syringe 48 2.79 200.1 200.3 99.9 1.4
7 Syringe 45 3.07 258.0 247.6 104.2 1.2
8 Syringe 48 4.53 382.1 379.2 100.8 1.2
9 Syringe 48 5.67 657.9 651.1 101.0 0.9
1Micro-
capillary
12 0.58 28.0 28.5 98.3 2.1
2Micro-
capillary
12 0.81 51.3 49.9 102.7 1.6
3Micro-
capillary
12 1.10 95.6 85.6 111.7 1.2
4Micro-
capillary
12 0.67 106.0 107.0 99.0 0.6
Table E-19 Rapidlab 1245 Recovery and Precision Testing—pCO
2
Level Mode n WRSD
*
Observed Expected Recovery WRCV