E-26 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
The following tables summarize the results of the Rapidlab 1245 CO-ox module recovery
and precision testing.
Table E-21 Rapidlab 1245 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~6 g/dL and Oxyhemoglobin > 75%
Analyte Mode
*
* tHb mode performance is equivalent to sample device used
n
Mean
Reference
Mean
Bias
Pooled
RSD
WRSD: within-run standard deviation
tHb Capillary 60 6.18 0.30 0.09
tHb tHb 54 6.18 0.19 0.13
tHb Syringe 53 6.14 0.25 0.05
FO
2
Hb Capillary 60 87.99 -1.37 0.26
FO
2
Hb tHb 55 87.34 -0.80 0.22
FO
2
Hb Syringe 54 88.15 -0.85 0.22
FCOHb Capillary 60 5.17 0.50 0.20
FCOHb tHb 55 4.86 0.63 0.19
FCOHb Syringe 54 5.34 0.70 0.25
FMetHb Capillary 60 4.99 0.22 0.10
FMetHb tHb 55 5.50 -0.05 0.11
FMetHb Syringe 54 5.84 -0.03 0.10
FHHb tHb 55 2.30 0.02 0.13
FHHb Capillary 60 1.86 0.16 0.19
FHHb Syringe 54 0.67 -0.06 0.12
Table E-22 Rapidlab 1245 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~14 g/dL and Oxyhemoglobin > 75%
Analyte Mode
*
n
Mean
Reference
Mean
Bias
Pooled
RSD
tHb Capillary 60 14.12 0.38 0.10
tHb tHb 58 14.17 0.05 0.20
tHb Syringe 51 14.07 0.29 0.09
FO
2
Hb Capillary 60 87.32 -0.65 0.17
FO
2
Hb tHb 58 86.86 -0.20 0.24
FO
2
Hb Syringe 53 87.73 0.05 0.17
FCOHb Capillary 60 5.45 0.04 0.13
FCOHb tHb 58 5.09 0.24 0.22
FCOHb Syringe 53 5.19 0.41 0.14