EasyManua.ls Logo

Siemens Rapidlab 1200 - Page 387

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rapidlab 1200 Operator’s Guide: System Specifications E-27
02087462 Rev.V
FMetHb Capillary 60 5.22 -0.02 0.06
FMetHb tHb 58 5.22 -0.13 0.08
FMetHb Syringe 53 6.27 -0.41 0.06
FHHb tHb 58 2.83 -0.09 0.19
FHHb Capillary 60 2.01 0.16 0.12
FHHb Syringe 52 0.82 -0.29 0.11
* tHb mode performance is equivalent to sample device used
WRSD: within-run standard deviation
Table E-23 Rapidlab 1245 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~20 g/dL and Oxyhemoglobin > 75%
Analyte Mode
*
n
Mean
Reference
Mean
Bias
Pooled
RSD
tHb Capillary 59 20.14 0.05 0.16
tHb tHb 52 20.10 -0.07 0.22
tHb Syringe 52 20.23 -0.08 0.15
FO
2
Hb Capillary 61 86.04 0.61 0.27
FO
2
Hb tHb 53 86.73 1.63 0.35
FO
2
Hb Syringe 53 87.01 1.41 0.21
FCOHb Capillary 60 5.45 0.04 0.13
FCOHb tHb 58 5.09 0.24 0.22
FCOHb Syringe 53 5.19 0.41 0.14
FMetHb Capillary 60 5.10 -0.19 0.09
FMetHb tHb 52 5.80 -0.36 0.12
FMetHb Syringe 53 6.23 -0.38 0.15
FHHb tHb 54 1.01 -0.70 0.08
FHHb Capillary 61 2.98 -0.18 0.10
FHHb Syringe 53 0.84 -0.45 0.08
* tHb mode performance is equivalent to sample device used
WRSD: within-run standard deviation
Table E-22 Rapidlab 1245 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~14 g/dL and Oxyhemoglobin > 75%
Analyte Mode
*
n
Mean
Reference
Mean
Bias
Pooled
RSD

Table of Contents