Rapidlab 1200 Operator’s Guide: System Specifications E-35
02087462 Rev.V
1 pH, Glu,
Lac
Capillary
24 0.008 7.064 7.063 100.01 0.12
2 pH, Glu,
Lac
Capillary
24 0.005 7.235 7.228 100.09 0.06
3 pH, Glu,
Lac
Capillary
24 0.004 7.325 7.321 100.05 0.05
4 pH, Glu,
Lac
Capillary
24 0.002 7.431 7.427 100.06 0.03
5 pH, Glu,
Lac
Capillary
24 0.002 7.560 7.551 100.12 0.03
6 pH, Glu,
Lac
Capillary
24 0.004 7.648 7.649 99.98 0.05
1 pH, Glu,
Lac
Syringe
37 0.005 7.068 7.072 99.95 0.07
2 pH, Glu,
Lac
Syringe
48 0.004 7.237 7.234 100.03 0.06
3 pH, Glu,
Lac
Syringe
37 0.005 7.303 7.309 99.91 0.07
4 pH, Glu,
Lac
Syringe
47 0.002 7.438 7.439 99.99 0.02
5 pH, Glu,
Lac
Syringe
36 0.002 7.563 7.561 100.02 0.03
6 pH, Glu,
Lac
Syringe
48 0.002 7.651 7.650 100.01 0.02
* WRSD = within-run standard deviation
Table E-35 Rapidlab 1260 Recovery and Precision Testing—pCO
2
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 23 0.24 7.4 7.1 104.2 3.3
Table E-34 Rapidlab 1260 Recovery and Precision Testing—pH
Level Mode N WRSD
*
Observed Expected Recovery WRCV