EasyManua.ls Logo

Siemens Rapidlab 1200 - Page 405

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rapidlab 1200 Operator’s Guide: System Specifications E-45
02087462 Rev.V
3Micro-
capillary
12 0.08 5.7 6.7 85.6 1.3
4Micro-
capillary
9 0.14 9.9 11.8 83.3 1.4
5Micro-
capillary
12 0.52 13.1 16.2 80.7 4.0
1Micro-
syringe
24 0.08 1.2 1.2 101.0 6.2
2Micro-
syringe
24 0.08 3.2 3.6 86.9 2.6
3Micro-
syringe
23 0.17 5.8 6.4 91.8 2.9
4Micro-
syringe
23 0.36 9.3 10.9 85.7 3.8
5Micro-
syringe
24 0.15 11.7 13.9 84.1 1.3
1 pH, Glu,
Lac
Capillary
24 0.05 0.9 0.9 109.1 5.4
2 pH, Glu,
Lac
Capillary
24 0.04 2.8 2.8 99.0 1.6
3 pH, Glu,
Lac
Capillary
24 0.12 4.9 5.2 93.8 2.4
4 pH, Glu,
Lac
Capillary
24 0.20 11.2 12.3 91.4 1.7
5 pH, Glu,
Lac
Capillary
24 0.41 14.9 17.6 84.8 2.7
1 pH, Glu,
Lac
Syringe
45 0.02 0.9 0.8 105.0 2.4
2 pH, Glu,
Lac
Syringe
44 0.07 2.6 2.7 95.3 2.5
3 pH, Glu,
Lac
Syringe
48 0.09 5.7 6.0 94.5 1.6
Table E-42 Rapidlab 1260 Recovery and Precision Testing—Lactate
Level Mode N WRSD
*
Observed Expected Recovery WRCV

Table of Contents