E-44 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
4 pH, Glu,
Lac
Capillary
24 4.75 353 365 97.0 1.3
5 pH, Glu,
Lac
Capillary
23 14.82 500 525 95.2 3.0
1 pH, Glu,
Lac
Syringe
48 1.18 48 43 112.8 2.5
2 pH, Glu,
Lac
Syringe
48 2.01 93 87 107.0 2.2
3 pH, Glu,
Lac
Syringe
43 2.91 216 213 101.6 1.3
4 pH, Glu,
Lac
Syringe
48 4.93 377 382 98.6 1.3
5 pH, Glu,
Lac
Syringe
48 6.15 500 520 96.1 1.2
* WRSD = within-run standard deviation
Table E-42 Rapidlab 1260 Recovery and Precision Testing—Lactate
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 21 0.09 1.5 1.5 97.5 6.2
2 Capillary 24 0.06 2.1 2.2 96.8 3.0
3 Capillary 21 0.08 4.7 5.1 91.4 1.7
4 Capillary 24 0.16 9.1 11.0 83.2 1.8
5 Capillary 21 0.17 11.7 14.2 82.4 1.5
1 Syringe 45 0.03 1.1 1.3 89.8 3.0
2 Syringe 48 0.05 2.1 2.2 94.6 2.4
3 Syringe 45 0.15 5.1 5.6 90.5 3.1
4 Syringe 48 0.20 9.3 10.9 85.4 2.2
5 Syringe 45 0.21 12.1 14.3 84.3 1.7
1Micro-
capillary
12 0.09 1.1 1.0 104.4 8.1
2Micro-
capillary
9 0.09 4.3 4.4 98.4 2.2
Table E-41 Rapidlab 1260 Recovery and Precision Testing—Glucose
Level Mode N WRSD
*
Observed Expected Recovery WRCV