Rapidlab 1200 Operator’s Guide: System Specifications E-43
02087462 Rev.V
Table E-41 Rapidlab 1260 Recovery and Precision Testing—Glucose
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 24 0.76 50.2 46 109.1 1.5
2Capillary241.298988101.61.4
3 Capillary 24 3.93 216 211 102.4 1.8
4 Capillary 24 6.25 330 350 94.3 1.9
5 Capillary 24 5.75 454 466 97.4 1.3
1 Syringe 48 0.99 44 40 109.4 2.3
2 Syringe 48 1.83 86 82 105.2 2.1
3 Syringe 48 2.87 214 212 101.4 1.3
4 Syringe 48 3.52 338 357 94.6 1.0
5 Syringe 47 5.45 469 467 100.4 1.2
1Micro-
capillary
12 1.32 46 44 103.6 2.9
2Micro-
capillary
12 0.71 88 88 100.6 0.8
3Micro-
capillary
12 1.55 221 219 100.6 0.7
4Micro-
capillary
12 4.44 331 343 96.5 1.3
5Micro-
capillary
12 7.98 448 452 99.1 1.8
1Micro-
syringe
24 0.84 49 48 94.3 1.3
2Micro-
syringe
24 2.10 99 98 100.3 1.0
3Micro-
syringe
24 1.61 224 222 97.8 2.1
4Micro-
syringe
22 4.72 350 371 93.5 1.3
5Micro-
syringe
24 4.73 453 452 98.9 1.4
1 pH, Glu,
Lac
Capillary
21 1.50 50 46 109.4 3.0
2 pH, Glu,
Lac
Capillary
24 1.07 96 93 104.1 1.1
3 pH, Glu,
Lac
Capillary
22 3.98 205 205 100.0 1.9