E-42 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
Table E-40 Rapidlab 1260 Recovery and Precision Testing—Ca
++
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 24 0.015 0.58 0.58 98.5 2.5
2 Capillary 24 0.015 1.27 1.26 100.6 1.2
3 Capillary 25 0.006 1.35 1.36 99.5 0.4
4 Capillary 24 0.017 1.90 1.93 98.8 0.9
5 Capillary 24 0.047 2.30 2.36 97.8 2.0
6 Capillary 24 0.089 4.06 4.09 99.2 2.2
1 Syringe 48 0.017 0.57 0.58 97.8 2.9
2 Syringe 48 0.012 1.26 1.26 100.4 1.0
3 Syringe 46 0.009 1.34 1.36 98.5 0.6
4 Syringe 51 0.027 1.87 1.93 97.2 1.4
5 Syringe 45 0.046 2.30 2.36 97.6 2.0
6 Syringe 45 0.099 4.22 4.09 103.1 2.4
1Micro-
capillary
12 0.034 0.57 0.56 101.6 5.9
2Micro-
capillary
12 0.018 1.24 1.24 100.0 1.4
3Micro-
capillary
12 0.005 1.30 1.30 100.0 0.4
4Micro-
capillary
120.0241.961.9899.01.2
5Micro-
capillary
12 0.050 2.39 2.38 100.5 2.1
6Micro-
capillary
120.0553.893.9897.81.4
1Micro-
syringe
230.0110.540.5694.82.0
2Micro-
syringe
240.0151.231.2499.41.2
3Micro-
syringe
24 0.006 1.30 1.30 100.3 0.5
4Micro-
syringe
22 0.074 1.99 1.98 100.4 3.7
5Micro-
syringe
24 0.032 2.41 2.38 101.4 1.4
6Micro-
syringe
24 0.092 4.02 3.98 101.0 2.3
* WRSD = within-run standard deviation