Table E-39 1260 Recovery and Precision Testing—Cl
-
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1Capillary240.58727399.20.8
2Capillary240.29909199.30.3
3Capillary240.58979898.80.6
4 Capillary 24 0.58 104 106 97.8 0.6
5 Capillary 24 0.50 110 111 99.1 0.5
6 Capillary 24 1.08 146 148 98.8 0.7
1 Syringe 47 0.57 70 71 98.2 0.8
2 Syringe 48 0.41 90 91 99.1 0.5
3 Syringe 48 0.79 98 99 99.0 0.8
4 Syringe 48 0.80 104 105 99.0 0.8
5 Syringe 48 0.56 110 110 99.3 0.5
6 Syringe 43 1.39 145 146 99.1 1.0
1Micro-
capillary
12 0.81 81 80 100.7 1.0
2Micro-
capillary
12 0.31 98 98 100.1 0.3
3Micro-
capillary
12 0.61 99 100 99.3 0.6
4Micro-
capillary
12 1.14 104 105 99.1 1.1
5Micro-
capillary
12 0.53 115 116 99.2 0.5
6Micro-
capillary
12 3.04 142 141 100.5 2.1
1Micro-
syringe
24 0.48 81 81 100.6 0.6
2Micro-
syringe
24 0.31 95 96 99.8 0.3
3Micro-
syringe
23 0.22 98 99 98.8 0.2
4Micro-
syringe
24 0.75 105 106 99.6 0.7
5Micro-
syringe
24 0.84 116 116 100.2 0.7
6Micro-
syringe
22 1.30 144 143 100.4 0.9
* WRSD = within-run standard deviation