E-40 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
Table E-38 Rapidlab 1260 Recovery and Precision Testing—K
+
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 21 0.096 1.21 1.18 102.2 8.0
2 Capillary 24 0.046 2.56 2.63 97.3 1.8
3 Capillary 24 0.034 4.19 4.12 101.7 0.8
4 Capillary 23 0.107 6.05 5.74 105.2 1.8
5 Capillary 21 0.084 7.88 7.99 98.6 1.1
6 Capillary 23 0.163 11.55 11.31 102.1 1.4
1 Syringe 45 0.078 1.19 1.18 100.5 6.5
2 Syringe 45 0.024 2.64 2.63 100.1 0.9
3 Syringe 45 0.026 4.12 4.12 100.1 0.6
4 Syringe 45 0.052 5.70 5.74 99.2 0.9
5 Syringe 45 0.116 8.18 7.99 102.3 1.4
6 Syringe 45 0.118 11.69 11.31 103.4 1.0
1Micro-
capillary
12 0.078 1.45 1.15 126.1 5.3
2Micro-
capillary
13 0.027 2.84 2.82 100.8 1.0
3Micro-
capillary
120.0243.723.7798.40.7
4Micro-
capillary
120.0425.575.7197.60.7
5Micro-
capillary
110.0737.588.0594.11.0
6Micro-
capillary
12 0.190 11.86 11.97 99.1 1.6
1Micro-
syringe
22 0.088 1.29 1.15 112.5 6.8
2Micro-
syringe
24 0.031 2.89 2.82 102.6 1.1
3Micro-
syringe
240.0193.733.7798.70.5
4Micro-
syringe
240.0445.615.7198.20.8
5Micro-
syringe
200.1597.978.0599.02.0
6Micro-
syringe
18 0.269 11.90 11.97 99.4 2.3
* WRSD = within-run standard deviation