Rapidlab 1200 Operator’s Guide: System Specifications E-39
02087462 Rev.V
Table E-37 Rapidlab 1260 Recovery and Precision Testing—Na
+
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 24 0.26 113.8 111.2 102.3 0.2
2 Capillary 24 0.21 130.6 129.4 100.9 0.2
3 Capillary 24 0.30 135.0 134.4 100.4 0.2
4 Capillary 24 0.55 142.8 141.5 100.9 0.4
5 Capillary 24 0.43 147.9 148.6 99.5 0.3
6 Capillary 24 0.85 175.7 179.8 97.7 0.5
1 Syringe 47 0.27 114.3 111.2 102.7 0.2
2 Syringe 48 0.29 129.7 129.4 100.2 0.2
3 Syringe 48 0.34 134.8 134.4 100.3 0.3
4 Syringe 48 0.71 141.2 141.5 99.8 0.5
5 Syringe 48 0.34 148.9 148.6 100.2 0.2
6 Syringe 48 0.78 174.6 179.8 97.1 0.4
1Micro-
capillary
12 0.62 110.2 109.9 100.3 0.6
2Micro-
capillary
12 0.18 129.5 130.1 99.6 0.1
3Micro-
capillary
12 0.60 131.7 132.0 99.8 0.5
4Micro-
capillary
12 1.11 143.1 143.8 99.5 0.8
5Micro-
capillary
12 0.29 149.6 150.8 99.2 0.2
6Micro-
capillary
12 1.29 175.4 178.1 98.5 0.7
1Micro-
syringe
24 0.41 109.6 109.9 99.8 0.4
2Micro-
syringe
24 0.15 130.6 130.1 100.4 0.1
3Micro-
syringe
23 0.21 132.1 132.0 100.1 0.2
4Micro-
syringe
24 0.62 143.5 143.8 99.8 0.4
5Micro-
syringe
24 0.47 150.1 150.8 99.5 0.3
6Micro-
syringe
22 0.99 178.7 178.1 100.3 0.6
* WRSD = within-run standard deviation