E-38 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
9 Syringe 42 13.76 652.3 648.5 100.6 2.1
1Micro-
capillary
12 0.20 26.2 28.4 92.3 0.8
2Micro-
capillary
12 0.57 46.6 49.8 93.5 1.2
3Micro-
capillary
12 0.61 83.1 85.9 96.7 0.7
4Micro-
capillary
12 0.89 105.0 106.7 98.5 0.9
5Micro-
capillary
12 0.75 149.3 150.4 99.3 0.5
6Micro-
capillary
12 5.00 201.5 198.9 101.3 2.5
7Micro-
capillary
12 2.22 263.9 250.6 105.3 0.8
8Micro-
capillary
12 3.91 380.4 376.6 101.0 1.0
9Micro-
capillary
12 8.65 656.9 647.0 101.5 1.3
1Micro-
syringe
24 0.26 26.5 28.6 92.8 1.0
2Micro-
syringe
24 0.37 46.6 49.5 94.2 0.8
3Micro-
syringe
24 0.73 81.3 85.6 95.1 0.9
4Micro-
syringe
23 0.88 104.1 106.1 98.1 0.8
5Micro-
syringe
24 1.33 147.8 149.7 98.7 0.9
6Micro-
syringe
24 3.36 201.4 199.9 100.7 1.7
7Micro-
syringe
24 2.75 267.9 249.6 107.4 1.0
8Micro-
syringe
24 4.02 386.6 378.4 102.2 1.0
9Micro-
syringe
24 12.58 670.7 643.4 104.2 1.9
* WRSD = within-run standard deviation
Table E-36 Rapidlab 1260 Recovery and Precision Testing—pO
2
Level Mode N WRSD
*
Observed Expected Recovery WRCV