Rapidlab 1200 Operator’s Guide: System Specifications E-37
02087462 Rev.V
2Micro-
syringe
24 0.56 20.7 21.4 96.7 2.7
3Micro-
syringe
24 0.50 28.2 28.6 98.6 1.8
4Micro-
syringe
24 1.00 34.7 35.7 97.4 2.9
5Micro-
syringe
24 1.31 48.2 49.5 97.4 2.7
6Micro-
syringe
24 1.49 72.2 71.4 101.1 2.1
7Micro-
syringe
24 2.67 97.1 99.8 97.3 2.8
8Micro-
syringe
23 2.02 131.4 134.4 97.8 1.5
9Micro-
syringe
24 4.15 229.7 228.2 100.7 1.8
* WRSD = within-run standard deviation
Table E-36 Rapidlab 1260 Recovery and Precision Testing—pO
2
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 21 0.31 30.4 28.4 106.9 1.0
2 Capillary 21 0.36 49.6 49.9 99.5 0.7
3 Capillary 24 0.66 89.7 84.9 105.6 0.7
4 Capillary 21 0.51 105.2 106.8 98.5 0.5
5 Capillary 24 1.52 150.8 148.6 101.5 1.0
6 Capillary 21 0.60 204.7 198.9 102.9 0.3
7 Capillary 24 1.16 255.4 247.6 103.1 0.5
8 Capillary 21 1.74 379.5 376.5 100.8 0.5
9 Capillary 20 6.79 654.4 648.1 101.0 1.0
1 Syringe 45 0.52 29.8 28.6 104.3 1.8
2 Syringe 42 0.35 49.2 49.9 98.7 0.7
3 Syringe 48 0.75 87.6 85.1 102.9 0.9
4 Syringe 42 0.67 106.4 106.9 99.5 0.6
5 Syringe 48 2.18 148.5 148.9 99.7 1.5
6 Syringe 45 2.21 200.3 200.0 100.1 1.1
7 Syringe 47 11.16 256.6 248.3 103.3 4.3
8 Syringe 45 7.47 371.0 378.6 98.0 2.0
Table E-35 Rapidlab 1260 Recovery and Precision Testing—pCO
2
Level Mode N WRSD
*
Observed Expected Recovery WRCV