E-58 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
1Micro-
syringe
24 0.41 28.0 28.6 98.0 1.5
2Micro-
syringe
24 0.22 48.3 49.5 97.5 0.5
3Micro-
syringe
24 0.97 84.1 85.6 98.3 1.1
4Micro-
syringe
24 1.53 105.5 106.1 99.5 1.5
5Micro-
syringe
24 1.17 149.9 149.7 100.1 0.8
6Micro-
syringe
24 2.52 200.5 199.9 100.3 1.3
7Micro-
syringe
24 2.74 263.0 249.6 105.4 1.0
8Micro-
syringe
24 4.31 376.9 378.4 99.6 1.1
9Micro-
syringe
24 8.16 640.8 643.4 99.6 1.3
* WRSD = within-run standard deviation
Table E-57 Rapidlab 1265 Recovery and Precision Testing—Na
+
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 21 0.18 113.6 111.2 102.1 0.2
2 Capillary 21 0.19 129.6 129.4 100.1 0.1
3 Capillary 21 0.42 134.0 134.4 99.7 0.3
4 Capillary 21 0.84 142.1 141.5 100.4 0.6
5 Capillary 21 0.25 147.4 148.6 99.2 0.2
6 Capillary 21 1.06 175.3 179.8 97.5 0.6
1 Syringe 45 0.41 113.5 111.2 102.1 0.4
2 Syringe 45 0.40 129.0 129.4 99.6 0.3
3 Syringe 45 0.36 133.6 134.4 99.4 0.3
4 Syringe 43 0.33 140.1 141.5 99.0 0.2
5 Syringe 42 0.25 148.5 148.6 99.9 0.2
6 Syringe 45 0.65 173.7 179.8 96.6 0.4
1Micro-
capillary
12 0.56 110.0 109.9 100.1 0.5
2Micro-
capillary
12 0.27 129.9 130.1 99.9 0.2
Table E-56 Rapidlab 1265 Recovery and Precision Testing—pO
2
Level Mode N WRSD
*
Observed Expected Recovery WRCV