Rapidlab 1200 Operator’s Guide: System Specifications E-59
02087462 Rev.V
3Micro-
capillary
12 0.11 133.3 132.0 101.0 0.1
4Micro-
capillary
12 0.60 144.0 143.8 100.1 0.4
5Micro-
capillary
12 0.40 150.4 150.8 99.7 0.3
6Micro-
capillary
12 0.83 177.6 178.1 99.7 0.5
1Micro-
syringe
24 0.71 109.7 109.9 99.8 0.6
2Micro-
syringe
24 0.30 130.4 130.1 100.2 0.2
3Micro-
syringe
24 0.35 132.2 132.0 100.2 0.3
4Micro-
syringe
24 0.44 144.1 143.8 100.2 0.3
5Micro-
syringe
24 0.68 151.2 150.8 100.3 0.5
6Micro-
syringe
24 1.09 179.5 178.1 100.8 0.6
* WRSD = within-run standard deviation
Table E-58 Rapidlab 1265 Recovery and Precision Testing—K
+
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 24 0.043 1.10 1.18 92.8 3.9
2 Capillary 21 0.028 2.54 2.63 96.5 1.1
3 Capillary 21 0.046 4.16 4.12 101.0 1.1
4 Capillary 21 0.036 6.04 5.74 105.2 0.6
5 Capillary 24 0.120 7.99 7.99 100.0 1.5
6 Capillary 21 0.115 11.48 11.31 101.5 1.0
1 Syringe 45 0.113 1.10 1.18 92.7 10.3
2 Syringe 45 0.027 2.63 2.63 99.9 1.0
3 Syringe 45 0.010 4.08 4.12 98.9 0.3
4 Syringe 45 0.065 5.65 5.74 98.4 1.2
5 Syringe 45 0.108 8.24 7.99 103.0 1.3
6 Syringe 45 0.153 11.87 11.31 104.9 1.3
Table E-57 Rapidlab 1265 Recovery and Precision Testing—Na
+
Level Mode N WRSD
*
Observed Expected Recovery WRCV