E-60 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
1Micro-
capillary
11 0.063 1.27 1.15 110.4 4.9
2Micro-
capillary
12 0.035 2.77 2.82 98.0 1.3
3Micro-
capillary
12 0.066 3.85 3.77 101.9 1.7
4Micro-
capillary
12 0.034 5.63 5.71 98.6 0.6
5Micro-
capillary
12 0.085 7.72 8.05 95.9 1.1
6Micro-
capillary
11 0.242 12.28 11.97 102.6 2.0
1Micro-
syringe
25 0.091 1.15 1.15 99.6 7.9
2Micro-
syringe
24 0.034 2.78 2.82 98.7 1.2
3Micro-
syringe
24 0.016 3.77 3.77 99.8 0.4
4Micro-
syringe
24 0.058 5.71 5.71 100.1 1.0
5Micro-
syringe
24 0.134 8.22 8.05 102.1 1.6
6Micro-
syringe
24 0.280 12.38 11.97 103.4 2.3
* WRSD = within-run standard deviation
Table E-59 Rapidlab 1265 Recovery and Precision Testing—Cl
-
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 21 0.38 72 73 99.2 0.5
2 Capillary 21 0.38 90 91 99.6 0.5
3 Capillary 21 0.38 97 98 99.3 0.3
4 Capillary 21 0.38 105 106 98.5 0.6
5 Capillary 21 0.38 111 111 99.8 0.4
6 Capillary 20 0.38 148 148 100.0 1.4
1 Syringe 45 0.38 70 71 98.0 1.4
2 Syringe 45 0.38 90 91 99.7 0.5
3 Syringe 45 0.42 99 99 100.2 0.4
4 Syringe 43 0.69 106 105 100.3 0.7
Table E-58 Rapidlab 1265 Recovery and Precision Testing—K
+
Level Mode N WRSD
*
Observed Expected Recovery WRCV