Rapidlab 1200 Operator’s Guide: System Specifications E-61
02087462 Rev.V
5 Syringe 42 0.38 110 111 99.7 0.3
6 Syringe 45 0.98 146 146 100.6 0.7
1Micro-
capillary
12 0.68 80 80 100.1 0.9
2Micro-
capillary
12 0.43 98 98 100.1 0.4
3Micro-
capillary
12 0.43 98 100 98.8 0.4
4Micro-
capillary
12 0.53 104 105 98.8 0.5
5Micro-
capillary
11 0.31 116 117 99.5 0.3
6Micro-
capillary
12 1.22 141 141 99.3 0.9
1Micro-
syringe
24 0.53 81 81 100.1 0.7
2Micro-
syringe
24 0.31 95 96 99.5 0.3
3Micro-
syringe
24 0.37 98 99 98.8 0.4
4Micro-
syringe
24 0.78 106 106 99.9 0.7
5Micro-
syringe
24 0.65 117 116 101.0 0.6
6Micro-
syringe
24 1.63 145 144 101.1 1.1
* WRSD = within-run standard deviation
Table E-60 Rapidlab 1265 Recovery and Precision Testing—Ca
++
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 18 0.009 0.57 0.58 97.0 1.6
2 Capillary 21 0.015 1.26 1.26 100.2 1.2
3 Capillary 18 0.004 1.34 1.36 98.4 0.3
4 Capillary 21 0.023 1.90 1.93 98.7 1.2
5 Capillary 18 0.026 2.29 2.36 97.1 1.1
6 Capillary 21 0.039 4.08 4.09 99.6 1.0
1 Syringe 40 0.011 0.55 0.58 94.6 2.0
2 Syringe 45 0.009 1.25 1.26 99.3 0.7
Table E-59 Rapidlab 1265 Recovery and Precision Testing—Cl
-
Level Mode N WRSD
*
Observed Expected Recovery WRCV